A novel cell count method, which can improve the count efficiency and reduce contamination problem, was presented using micro lattice engraved on culture dish. The micro lattice has feature of $50{\mu}m{\times}50{\mu}m$ rectangular shape, 2㎛ line width, and 2㎛ depth in $3mm{\times}3mm$ area. In this paper, nickel mold was fabricated with thickness of 3mm and diameter of 80 mm, and transcription of the micro lattice on a polystyrene cell culture dish was performed by hot embossing at $200\;^{\circ}C$. The tedious and error-prone harvest/load processes of conventional cell counts with a hemocytometer could be omitted, and these advantages became magnified during periodical counts involving long-term cultures. SupT1 cells and HeLa cells were cultivated with the dish for 7 days in $CO2 incubator and counted as $371.84/mm^2$ and $123.36/mm^2$, respectively, during the cultivation without harmful effects on the cells.