Ramp-Preserving Denoising for Conductivity Image Reconstruction in Magnetic Resonance Electrical Impedance Tomography

Cited 18 time in webofscience Cited 0 time in scopus
  • Hit : 730
  • Download : 0
DC FieldValueLanguage
dc.contributor.authorLee, Chang-Ockko
dc.contributor.authorJeon, Kiwanko
dc.contributor.authorAhn, Seonminko
dc.contributor.authorKim, Hyung Joongko
dc.contributor.authorWoo, Eung Jeko
dc.date.accessioned2013-03-11T06:41:10Z-
dc.date.available2013-03-11T06:41:10Z-
dc.date.created2012-02-06-
dc.date.created2012-02-06-
dc.date.created2012-02-06-
dc.date.issued2011-07-
dc.identifier.citationIEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, v.58, no.7, pp.2038 - 2050-
dc.identifier.issn0018-9294-
dc.identifier.urihttp://hdl.handle.net/10203/98543-
dc.description.abstractIn magnetic resonance electrical impedance tomography, among several conductivity image reconstruction algorithms, the harmonic B(z) algorithm has been successfully applied to Bz data from phantoms and animals. The algorithm is, however, sensitive to measurement noise in B(z) data. Especially, in in vivo animal and human experiments where injection current amplitudes are limited within a few milliampere at most, measured B(z) data tend to have a low SNR. In addition, magnetic resonance (MR) signal void in outer layers of bones and gas-filled organs, for example, produces salt-pepper noise in the MR phase and, consequently, B(z) images. The B(z) images typically present areas of sloped transitions, which can be assimilated to ramps. Conductivity contrasts change ramp slopes in B(z) images and it is critical to preserve positions of those ramps to correctly recover edges in conductivity images. In this paper, we propose a ramp-preserving denoising method utilizing a structure tensor. Using an eigenvalue analysis, we identified local regions of salt-pepper noise. Outside the identified local regions, we applied an anisotropic smoothing to reduce noise while preserving their ramp structures. Inside the local regions of salt-pepper noise, we used an isotropic smoothing. After validating the proposed denoising method through numerical simulations, we applied it to in vivo animal imaging experiments. Both numerical simulation and experimental results show significant improvements in the quality of reconstructed conductivity images.-
dc.languageEnglish-
dc.publisherIEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC-
dc.titleRamp-Preserving Denoising for Conductivity Image Reconstruction in Magnetic Resonance Electrical Impedance Tomography-
dc.typeArticle-
dc.identifier.wosid000291890000019-
dc.identifier.scopusid2-s2.0-79959572012-
dc.type.rimsART-
dc.citation.volume58-
dc.citation.issue7-
dc.citation.beginningpage2038-
dc.citation.endingpage2050-
dc.citation.publicationnameIEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING-
dc.contributor.localauthorLee, Chang-Ock-
dc.contributor.nonIdAuthorJeon, Kiwan-
dc.contributor.nonIdAuthorAhn, Seonmin-
dc.contributor.nonIdAuthorKim, Hyung Joong-
dc.contributor.nonIdAuthorWoo, Eung Je-
dc.type.journalArticleArticle-
dc.subject.keywordAuthorConductivity image-
dc.subject.keywordAuthormagnetic resonance electrical impedance tomography (MREIT)-
dc.subject.keywordAuthorpartial differential equation (PDE)-based denoising-
dc.subject.keywordAuthorramp-preserving denoising-
dc.subject.keywordPlusMREIT SYSTEM-
dc.subject.keywordPlusMR-EIT-
dc.subject.keywordPlusDENSITY-
dc.subject.keywordPlusALGORITHM-
dc.subject.keywordPlusSIMULATION-
dc.subject.keywordPlusDIFFUSION-
Appears in Collection
MA-Journal Papers(저널논문)
Files in This Item
There are no files associated with this item.
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 18 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0