Development of in vivo Confocal Microscope Combined with Reflection and Fluorescence Imaging Modes

Cited 1 time in webofscience Cited 0 time in scopus
  • Hit : 277
  • Download : 0
We propose a new in vivo confocal microscope capable of reflection and fluorescence imaging simultaneously with the use of 488nm, 660nm, and 830nm lasers. We measured the lateral resolution of less than 0.8m, the axial resolution of less than 3.5m, the focal shift of 1.25m, and the imaging depth up to 210m. Ex vivo and in vivo images are acquired from bovine artery endothelial cell, mouse, and human skin tissues. Two imaging modes can provide potentials that the accuracy of diagnosis can be improved from different point of views.