THE EFFECT OF SURFACE ROUGHNESS OF CSI(TL) MICRO-COLUMNS ON THE RESOLUTION OF THE X-RAY IMAGE; OPTICAL SIMULATION STUDY

Cited 0 time in webofscience Cited 0 time in scopus
  • Hit : 500
  • Download : 0
Micro-columnar CsI(Tl) is the most popular scintillator material which is used for many indirect digital X-ray imaging detectors. The light scattering at the surface of micro-columnar CsI(Tl) scintillator was studied to find the correlation between the surface roughness and the resultant image resolution of indirect X-ray imaging detectors. Using a commercially available optical simulation program, Light Tools, MTF (Modulation Transfer Function) curves of the CsI(Tl) film thermally evaporated on glass substrate with different thickness were calculated and compared with the experimental estimation of MTF values by the edge X-ray image method and CCD camera. It was found that the standard deviation value of Gaussian scattering model which is determined by the surface roughness of micro-columns could certainly change the MTF value of image sensors. This model and calculation methodology will be beneficial to estimate the overall performance of indirect X-ray imaging system with CsI(Tl) scintillator film for optimum design depending on its application.
Publisher
대한방사선방어학회
Issue Date
2009-03
Language
English
Citation

방사선방어학회지, v.34, no.1, pp.25 - 30

ISSN
0253-4231
URI
http://hdl.handle.net/10203/93492
Appears in Collection
NE-Journal Papers(저널논문)
Files in This Item
There are no files associated with this item.

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0