DC Field | Value | Language |
---|---|---|
dc.contributor.author | 이소연 | ko |
dc.contributor.author | 이동우 | ko |
dc.contributor.author | 조영호 | ko |
dc.date.accessioned | 2013-03-08T03:42:26Z | - |
dc.date.available | 2013-03-08T03:42:26Z | - |
dc.date.created | 2012-02-06 | - |
dc.date.created | 2012-02-06 | - |
dc.date.issued | 2005-12 | - |
dc.identifier.citation | 대한기계학회논문집 A, v.29, no.12, pp.1615 - 1620 | - |
dc.identifier.issn | 1226-4873 | - |
dc.identifier.uri | http://hdl.handle.net/10203/92040 | - |
dc.description.abstract | We present a novel flow-rate independent cell counter using a fixed control volume between double electrical sensing zones. The previous device based on the single electrical cell sensing in a given flow-rate requires an accurate fluid volume measurement or precision flow rate control. The present cell counter, however, offers the flow-rate independent method for the cell concentration measurement with counting cells in a fixed control volume of . In the experimental study, using the RBC (Red Blood Cell), we have compared the measured RBC concentrations from the fabricated devices with those from Hemacytometer. The previous and present devices show the maximum errors of , which are in the measurement error range of Hemacytometer (about ). The present device also shows the flow-rate independent performance at the constant flow-rates ( and ) and the varying flow-rate (4, 2, and ). Therefore, we demonstrate that the present cell counter is a simple and automated method for the cell concentration measurement without requiring an accurate fluid measurement and precision flow-rate control. | - |
dc.language | Korean | - |
dc.publisher | 대한기계학회 | - |
dc.title | 다수의 계수구역간의 검사체적을 이용한 소형 세포농도센서 | - |
dc.title.alternative | Micro Cell Counter using a Fixed Control Volume between Double Electrical Sensing Zones | - |
dc.type | Article | - |
dc.type.rims | ART | - |
dc.citation.volume | 29 | - |
dc.citation.issue | 12 | - |
dc.citation.beginningpage | 1615 | - |
dc.citation.endingpage | 1620 | - |
dc.citation.publicationname | 대한기계학회논문집 A | - |
dc.identifier.kciid | ART000976623 | - |
dc.contributor.localauthor | 조영호 | - |
dc.contributor.nonIdAuthor | 이소연 | - |
dc.contributor.nonIdAuthor | 이동우 | - |
dc.subject.keywordAuthor | Cell Counter | - |
dc.subject.keywordAuthor | Coulter Counter | - |
dc.subject.keywordAuthor | Electrical Sensing Zone | - |
dc.subject.keywordAuthor | Cell Concentration Sensor | - |
dc.subject.keywordAuthor | 세포 계수기 | - |
dc.subject.keywordAuthor | 컬터 카운터 | - |
dc.subject.keywordAuthor | 전기적 계수구역 | - |
dc.subject.keywordAuthor | 세포 농도측정기 | - |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.