A miniaturized time-of-flight mass spectrometer with an electron impact ionization source and sheet membrane introduction has been developed. The advantages and features of this mass spectrometer include high sensitivity, simple structure, low cost, compact volume with field portability, and ease of operation. A mass resolution of 400 at m/z 78 has been obtained with a 25 ern flight path length. Under optimized conditions, the detection limits for the volatile organic compounds (VOCs) studied were 0.2-10 ppm by volume with linear dynamic ranges greater than three orders of magnitude. The response times for various VOCs using a silicone membrane of 127 mu m thickness were in the range 4.5-20 s, which provides a sample analysis time of less than I minute. These results indicate that the membrane introduction/time-of-flight mass spectrometer will be useful for a wide range of field applications, particularly for environmental monitoring.
Publisher
KOREAN CHEMICAL SOC
Issue Date
2005-02
Language
English
Article Type
Article
Keywords

SEMIVOLATILE ORGANIC-COMPOUNDS; ION-TRAP; MULTIPHOTON IONIZATION; WATER; MINIATURE; VOLATILE; ARRAY; GAS; RESOLUTION; ANALYZERS

Citation

BULLETIN OF THE KOREAN CHEMICAL SOCIETY, v.26, no.2, pp.303 - 308

ISSN
0253-2964
URI
http://hdl.handle.net/10203/90136
Appears in Collection
Files in This Item
There are no files associated with this item.
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 10 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0