Simultaneous analysis of Np-237 and Pu isotopes in environmental samples by ICP-SF-MS coupled with automated sequential injection system

Cited 77 time in webofscience Cited 77 time in scopus
  • Hit : 292
  • Download : 0
DC FieldValueLanguage
dc.contributor.authorKim, CSko
dc.contributor.authorKim, CKko
dc.contributor.authorLee, Kun Jaiko
dc.date.accessioned2013-03-04T08:41:08Z-
dc.date.available2013-03-04T08:41:08Z-
dc.date.created2012-02-06-
dc.date.created2012-02-06-
dc.date.issued2004-
dc.identifier.citationJOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, v.19, no.6, pp.743 - 750-
dc.identifier.issn0267-9477-
dc.identifier.urihttp://hdl.handle.net/10203/82204-
dc.description.abstractA simultaneous analytical method for Np-237, Pu-239 and Pu-240 has been developed using sector field inductively coupled plasma mass spectrometry (ICP-SF-MS) detection combined with an automated sequential injection ( SI) separation system. The chemical purification of these actinides was carried out on an actinide-specific extraction chromatographic resin, TEVA-Spec, in an automated SI system. The similar chemical behaviors of Np and Pu on TEVA-Spec was confirmed by recovery test after treating with ascorbic acid in 5 M HNO3. TEVA-Spec maintained its re-usability, giving high and consistent chemical recovery, and showed good selectivity for Np and Pu and a low memory effect during the consecutive analysis of nine replicate samples. The high chemical recovery (>90%), reliable accuracy and repeatability of Np-237 and Pu isotope purification using the SI system were proved using standard reference materials. Low detection limits of 2.5, 2.1 and 0.42 fg ml(-1) were obtained for Np-237, Pu-239 and Pu-240, respectively, which make it possible to analyze these nuclides in samples of soil and sediment as low as 1 g. With the exception of the preparation process required prior to loading a sample, all chemical purifications were carried out automatically by the programmed SI system, and were completed within 1 hour, which included measurement by ICP-SF-MS. This method was successfully applied to the analysis of the environmental levels of Np-237 and Pu in soil, sediment and in biological samples.-
dc.languageEnglish-
dc.publisherROYAL SOC CHEMISTRY-
dc.subjectPLASMA-MASS-SPECTROMETRY-
dc.subjectEXTRACTION CHROMATOGRAPHIC-SEPARATIONS-
dc.subjectPLUTONIUM ISOTOPES-
dc.subjectACTINIDE ELEMENTS-
dc.subjectATOM RATIOS-
dc.subjectONLINE-
dc.subjectPRECONCENTRATION-
dc.subjectSEDIMENTS-
dc.subjectPHASE-
dc.subjectSOIL-
dc.titleSimultaneous analysis of Np-237 and Pu isotopes in environmental samples by ICP-SF-MS coupled with automated sequential injection system-
dc.typeArticle-
dc.identifier.wosid000221844800005-
dc.identifier.scopusid2-s2.0-3042589554-
dc.type.rimsART-
dc.citation.volume19-
dc.citation.issue6-
dc.citation.beginningpage743-
dc.citation.endingpage750-
dc.citation.publicationnameJOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY-
dc.identifier.doi10.1039/b400034j-
dc.contributor.localauthorLee, Kun Jai-
dc.contributor.nonIdAuthorKim, CS-
dc.contributor.nonIdAuthorKim, CK-
dc.type.journalArticleArticle-
dc.subject.keywordPlusPLASMA-MASS-SPECTROMETRY-
dc.subject.keywordPlusEXTRACTION CHROMATOGRAPHIC-SEPARATIONS-
dc.subject.keywordPlusPLUTONIUM ISOTOPES-
dc.subject.keywordPlusACTINIDE ELEMENTS-
dc.subject.keywordPlusATOM RATIOS-
dc.subject.keywordPlusONLINE-
dc.subject.keywordPlusPRECONCENTRATION-
dc.subject.keywordPlusSEDIMENTS-
dc.subject.keywordPlusPHASE-
dc.subject.keywordPlusSOIL-
Appears in Collection
NE-Journal Papers(저널논문)
Files in This Item
There are no files associated with this item.
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 77 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0