Simultaneous acquisition of arterial and venous angiograms by time-multiplexed dual slice imaging시분할 두 영역 영상에 의한 동맥과 정맥 혈관의 동시 영상법

Cited 0 time in webofscience Cited 0 time in scopus
  • Hit : 372
  • Download : 0
A 3-D angiography technique for producing separate artery and vein angiograms is proposed. By inserting a common saturation region between the artery and vein imaging slices, only incident blood flows on both sides were separately observable. Signals from artery and vein are acquired in time multiplexed manner in a given repetition time, thereby angiograms of the both artery and vein are obtained by one sequence. Using he proposed method, a set of separate 3-D artery and vein angiogram is obtained using KAIS 2.0 Tesla whole body MR imaging system. The imaging volume was 220×220×120㎣ with a resolution of 1.1mm(x)×0.86mm(y)×1.2mm(z) and total imaging time was 16.5 minutes.
Advisors
Cho, Zang-Heeresearcher조장희researcher
Description
한국과학기술원 : 전기 및 전자공학과,
Publisher
한국과학기술원
Issue Date
1990
Identifier
67375/325007 / 000881433
Language
eng
Description

학위논문(석사) - 한국과학기술원 : 전기 및 전자공학과, 1990.2, [ [iii], 55 p. ]

URI
http://hdl.handle.net/10203/39164
Link
http://library.kaist.ac.kr/search/detail/view.do?bibCtrlNo=67375&flag=dissertation
Appears in Collection
EE-Theses_Master(석사논문)
Files in This Item
There are no files associated with this item.

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0