Optimization of X-ray target parameters for a high-brightness microfocus X-ray tube

Cited 26 time in webofscience Cited 30 time in scopus
  • Hit : 849
  • Download : 4
Parameters of a transmission-type X-ray target have been optimized for a microfocus X-ray tube with an emphasis on maximizing the intensity of X-rays generated from the target while reducing the X-ray focal spot size. Monte Carlo simulation code MCNPX was used for the calculation of X-ray intensity generated from the X-ray tube. The calculation shows that the target thickness plays a dominant role in the enhancement of X-ray intensity. The optimum thickness to get maximum X-ray intensity for a given beam current was derived from the calculation for a broad range of electron energies from 30 keV to 150 keV. In addition, stability of a transmission-type target against localized thermal loading, which is attributed to the electron beam impinging on a micron-sized target region, was also investigated by the calculation of the temperature distribution within the target. From the analysis, we present the maximum allowable electron-beam power loading for a stable operation of the transmission-type target at different incident electron beam energies. (c) 2007 Elsevier B.V. All rights reserved.
Publisher
ELSEVIER SCIENCE BV
Issue Date
2007-11
Language
English
Article Type
Article
Keywords

MONTE-CARLO-SIMULATION; SYSTEM; BEAM; CODE

Citation

NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, v.264, no.2, pp.371 - 377

ISSN
0168-583X
DOI
10.1016/j.nimb.2007.09.023
URI
http://hdl.handle.net/10203/3243
Appears in Collection
NE-Journal Papers(저널논문)
Files in This Item
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 26 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0