An Intra-Body Power Transfer System With >1-mW Power Delivered to the Load and 3.3-V DC Output at 160-cm of on-Body Distance

Cited 9 time in webofscience Cited 0 time in scopus
  • Hit : 92
  • Download : 0
DC FieldValueLanguage
dc.contributor.authorCho, Hyungjooko
dc.contributor.authorSuh, Ji-Hoonko
dc.contributor.authorKim, Chulko
dc.contributor.authorHa, Sohmyungko
dc.contributor.authorJe, Minkyuko
dc.date.accessioned2023-03-06T05:01:38Z-
dc.date.available2023-03-06T05:01:38Z-
dc.date.created2023-03-06-
dc.date.created2023-03-06-
dc.date.created2023-03-06-
dc.date.issued2022-10-
dc.identifier.citationIEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, v.16, no.5, pp.852 - 866-
dc.identifier.issn1932-4545-
dc.identifier.urihttp://hdl.handle.net/10203/305465-
dc.description.abstractThis paper presents an intra-body power transfer (IBPT) system that can deliver power greater than 1mW across an on-body distance of 160 cm. A system simulation model is built for the characterization of the channel and optimization of the power transfer. Our system analysis and experimental validation demonstrate that 1 MHz is an optimal carrier frequency for IBPT in terms of power delivered to the load (PDL) and power efficiency (PE). Prototype TX and RX boards were built, and an IC was fabricated in a 180-nm CMOS process for the RX. The proposed RX IC consists of a voltage doubler (VD) and a charge pump (CP) to obtain a sufficiently high voltage conversion ratio (VCR). Among various rectifier topologies, the VD is the optimal topology for the power receiver front-end because the parasitic ground coupling capacitances, which inevitably exist in the IBPT system, act as an inherent input-coupling capacitance for the VD. The implemented VD utilizes a dynamic V-TH compensation (DVC) for its diode components. Compared to the conventional static V-TH compensation (SVC), DVC in the VD reduces the reverse leakage current of the diode, thus maximizing the power conversion efficiency (PCE) and VCR. In addition, the PDL is enhanced by inserting an inductor on the TX board. It reduces the backward-path impedance without increasing the RX volume, boosting the PDL by up to 9.9 times compared to the PDL without the inductor insertion. The proposed IBPT system delivers up to 178.8 mu W of power at 11.7% of maximum power efficiency with 3.3-V DC output voltage and even 1.385 mW of power with the inductor insertion, supporting various biomedical wearable sensors, such as ECGsensor modules.-
dc.languageEnglish-
dc.publisherIEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC-
dc.titleAn Intra-Body Power Transfer System With >1-mW Power Delivered to the Load and 3.3-V DC Output at 160-cm of on-Body Distance-
dc.typeArticle-
dc.identifier.wosid000928191200016-
dc.identifier.scopusid2-s2.0-85135761759-
dc.type.rimsART-
dc.citation.volume16-
dc.citation.issue5-
dc.citation.beginningpage852-
dc.citation.endingpage866-
dc.citation.publicationnameIEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS-
dc.identifier.doi10.1109/TBCAS.2022.3194278-
dc.contributor.localauthorJe, Minkyu-
dc.contributor.nonIdAuthorKim, Chul-
dc.contributor.nonIdAuthorHa, Sohmyung-
dc.description.isOpenAccessN-
dc.type.journalArticleArticle-
dc.subject.keywordAuthorIntegrated circuit modeling-
dc.subject.keywordAuthorCouplings-
dc.subject.keywordAuthorCapacitance-
dc.subject.keywordAuthorWearable sensors-
dc.subject.keywordAuthorLoad modeling-
dc.subject.keywordAuthorImpedance-
dc.subject.keywordAuthorBatteries-
dc.subject.keywordAuthorBody-area network (BAN)-
dc.subject.keywordAuthorbody-channel model-
dc.subject.keywordAuthorbody-channel communication (BCC)-
dc.subject.keywordAuthorenergy harvesting (EH)-
dc.subject.keywordAuthorintra-body power transfer (IBPT)-
dc.subject.keywordAuthorpower delivered to the load (PDL)-
dc.subject.keywordAuthorpower efficiency (PE)-
dc.subject.keywordAuthorvoltage doubler-
dc.subject.keywordAuthorwireless power transfer (WPT)-
dc.subject.keywordAuthorwearable devices-
dc.subject.keywordAuthorAC-DC converter-
dc.subject.keywordPlusRECTIFIER-
dc.subject.keywordPlusTRANSMISSION-
dc.subject.keywordPlusMODULATION-
Appears in Collection
EE-Journal Papers(저널논문)
Files in This Item
There are no files associated with this item.
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 9 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0