Measurement of an analyte concentration in test solution by using Helmholtz resonator for biosensor applications

Cited 6 time in webofscience Cited 6 time in scopus
  • Hit : 505
  • Download : 169
In this paper, an indirect method of measuring an analyte concentration in a test solution using the resonant frequency change of a Helmholtz resonator is proposed, using a novel architecture of Helmholtz resonator filled with two kinds of fluids (fixed fluid and test solution). Since the analyte concentration yields changes of density and sound speed of the test solution, the resonant frequency of the proposed Helmholtz resonator is affected by the analyte concentration of the test solution. From this effect, the analyte concentration of the test solution can be measured by the spectrum of acoustic resonance of the Helmholtz resonator. The experiment was done using a 3D-printed Helmholtz resonator system with an acoustic power source and detectors, which is consistent with analytical results and showed that the analyte concentration can be measured with higher sensitivity compared to conventional cantilever-type sensors. As an example application, the possibility of measuring glucose concentration of human blood was demonstrated, showing higher sensitivity and relatively low frequency range compared to previous resonance based methods.
Publisher
MDPI
Issue Date
2019-03
Language
English
Article Type
Article
Citation

SENSORS, v.19, no.5, pp.1 - 9

ISSN
1424-8220
DOI
10.3390/s19051127
URI
http://hdl.handle.net/10203/251785
Appears in Collection
ME-Journal Papers(저널논문)
Files in This Item
sensors-19-01127.pdf(525.64 kB)Download
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 6 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0