Performance Measurements of the SAFIR Prototype Detector With the STiC ASIC Readout

Cited 0 time in webofscience Cited 11 time in scopus
  • Hit : 178
  • Download : 0
DC FieldValueLanguage
dc.contributor.authorAhnen, Maxko
dc.contributor.authorBecker, Robertko
dc.contributor.authorBuck, Alfredko
dc.contributor.authorCasella, Chiarako
dc.contributor.authorCommichau, Volkerko
dc.contributor.authordi Calafiori, Diogoko
dc.contributor.authorDissertori, Guntherko
dc.contributor.authorEleftheriou, Afroditiko
dc.contributor.authorFischer, Jannisko
dc.contributor.authorHoward, Alexander S.ko
dc.contributor.authorIto, Mikikoko
dc.contributor.authorKhateri, Parisako
dc.contributor.authorKim, Jisooko
dc.contributor.authorLustermann, Wernerko
dc.contributor.authorRitzer, Christianko
dc.contributor.authorRoser, Ulfko
dc.contributor.authorRudin, Markusko
dc.contributor.authorSolevi, Paolako
dc.contributor.authorTsoumpas, Charalamposko
dc.contributor.authorWarnock, Geoffreyko
dc.contributor.authorWeber, Brunoko
dc.contributor.authorWyss, Matthiasko
dc.contributor.authorZagozdzinska-Bochenek, Agnieszkako
dc.date.accessioned2019-02-20T05:15:47Z-
dc.date.available2019-02-20T05:15:47Z-
dc.date.created2019-02-07-
dc.date.created2019-02-07-
dc.date.created2019-02-07-
dc.date.created2019-02-07-
dc.date.issued2018-05-
dc.identifier.citationIEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, v.2, no.3, pp.250 - 258-
dc.identifier.issn2469-7311-
dc.identifier.urihttp://hdl.handle.net/10203/250422-
dc.description.abstractThe small animal fast insert for mRi (SAFIR) positron emission tomography insert was proposed for quantitative dynamic acquisition inside a preclinical 7T magnetic resonance imaging scanner to study kinetics of short-lived tracers. For this purpose, the SAFIR readout should be capable of handling high count rates and achieving excellent timing performance. We evaluated one of the available application specific integrated circuits (ASICs) for SiPM readout, namely SiPM timing chip (STiC) version 3.1. In this paper, we show the performances of the SAFIR PET detector with the STiC ASIC readout. The SAFIR PET detector consists of an 8 x 8 array of lutetium yttrium oxyorthosilicate 2.1 mmx 2.1 mmx 12 mm crystals coupled, with optical grease, to an 8 x 8 array of SiPMs with a 2.0 mmx 2.0 mm photo-sensitive area. Signals from the individual SiPM channels were digitized by the STiC ASIC. Hit's arrival time and time-over-threshold (TOT) were recorded into time stamps with 50.2-ps least significant bit. We obtained an average energy resolution of 18.5% full width at half maximum (FWHM) at 511-keV photopeak after TOT nonlinearity correction and an average coincidence resolving time resolution of 244-ps FWHM with time walk correction that satisfy our requirements specification on the detector performance.-
dc.languageEnglish-
dc.publisherIEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC-
dc.titlePerformance Measurements of the SAFIR Prototype Detector With the STiC ASIC Readout-
dc.typeArticle-
dc.identifier.scopusid2-s2.0-85074263197-
dc.type.rimsART-
dc.citation.volume2-
dc.citation.issue3-
dc.citation.beginningpage250-
dc.citation.endingpage258-
dc.citation.publicationnameIEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES-
dc.identifier.doi10.1109/TRPMS.2018.2797484-
dc.contributor.localauthorKim, Jisoo-
dc.contributor.nonIdAuthorAhnen, Max-
dc.contributor.nonIdAuthorBecker, Robert-
dc.contributor.nonIdAuthorBuck, Alfred-
dc.contributor.nonIdAuthorCasella, Chiara-
dc.contributor.nonIdAuthorCommichau, Volker-
dc.contributor.nonIdAuthordi Calafiori, Diogo-
dc.contributor.nonIdAuthorDissertori, Gunther-
dc.contributor.nonIdAuthorEleftheriou, Afroditi-
dc.contributor.nonIdAuthorFischer, Jannis-
dc.contributor.nonIdAuthorHoward, Alexander S.-
dc.contributor.nonIdAuthorIto, Mikiko-
dc.contributor.nonIdAuthorKhateri, Parisa-
dc.contributor.nonIdAuthorLustermann, Werner-
dc.contributor.nonIdAuthorRitzer, Christian-
dc.contributor.nonIdAuthorRoser, Ulf-
dc.contributor.nonIdAuthorRudin, Markus-
dc.contributor.nonIdAuthorSolevi, Paola-
dc.contributor.nonIdAuthorTsoumpas, Charalampos-
dc.contributor.nonIdAuthorWarnock, Geoffrey-
dc.contributor.nonIdAuthorWeber, Bruno-
dc.contributor.nonIdAuthorWyss, Matthias-
dc.contributor.nonIdAuthorZagozdzinska-Bochenek, Agnieszka-
dc.description.isOpenAccessN-
dc.type.journalArticleArticle-
dc.subject.keywordAuthorPerformance-
dc.subject.keywordAuthorphotodetector technology-
dc.subject.keywordAuthorSiPM-
dc.subject.keywordAuthorsmall animal fast insert for mRi (SAFIR)-
dc.subject.keywordAuthorSiPM timing chip (STiC) application specific integrated circuit (ASIC)-
dc.subject.keywordPlusPOSITRON-EMISSION-TOMOGRAPHY-
dc.subject.keywordPlusSIMULTANEOUS PET-
dc.subject.keywordPlusMRI-
dc.subject.keywordPlusPET/MRI-
Appears in Collection
RIMS Journal Papers
Files in This Item
There are no files associated with this item.

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0