Monolithic expandable IC based portable fNIRS system design확장이 가능한 단일 집적회로 기반의 휴대용 근적외선 뇌 영상장치 개발

Cited 0 time in webofscience Cited 0 time in scopus
  • Hit : 720
  • Download : 0
DC FieldValueLanguage
dc.contributor.advisorBae, Hyeon-Min-
dc.contributor.advisor배현민-
dc.contributor.authorChoi, Jong-Kwan-
dc.contributor.author최종관-
dc.date.accessioned2017-03-29T02:48:15Z-
dc.date.available2017-03-29T02:48:15Z-
dc.date.issued2016-
dc.identifier.urihttp://library.kaist.ac.kr/search/detail/view.do?bibCtrlNo=663191&flag=dissertationen_US
dc.identifier.urihttp://hdl.handle.net/10203/222320-
dc.description학위논문(박사) - 한국과학기술원 : 전기및전자공학부, 2016.8 ,[iv, 53 p :]-
dc.description.abstractA custom tranceiver IC is designed to implement a portable brain imaging system based on functional near-infrared-spectroscopy (fNIRS) and the way to select optimum wavelength pairs is analyzed based on wavelength dependent optical properties. The fNIRS IC generates multichannel time-divided spread-spectrum codes and drives light emitting devices in the transmitter (Tx) chain, and performs optimum filtering, quantization, and serialization in the receiver chain. A dual slope ADC subsequent to an operational transconductance amplifier-C (OTA-C) based matched filter shares a capacitor to save area while achieving optimum signal-to-noise ratio (SNR) in the brain channel. The receiver (Rx) chain including an off-chip TIA ensures sufficient electrical SNR irrespective of the brain region for the accurate extraction of hemodynamic response. The minimum detectable light power of the Rx chain is 400 fW. The output power of the Tx is made adjustable by controlling the occurrence rate and the power of the time-divided spread-spectrum code to reduce subject dependent measurement variations. The proposed fNIRS system uses 780nm and 850nm pair to extract hemodynamics from detected signals and the validity of used wavelength pair is proved by comparison of contrast-to-noise ratio (CNR) depends on the input SNR and hemodynamics changes. The proposed fNIRS system measures 3cm separated 48 brain regions simultaneously, and the experimental results clearly verify the validity of the proposed portable fNIRS system.-
dc.languageeng-
dc.publisher한국과학기술원-
dc.subjectfunctional near-infrared-spectroscopy (fNIRS)-
dc.subjectbrain optical imaging-
dc.subjecthemodynamics-
dc.subjectmatched filter-
dc.subjecthigh SNR-
dc.subjectcontrast-to-noise ratio-
dc.subjectCDMA-
dc.subjectemission power control-
dc.subjectportable-
dc.subjectreal-time-
dc.subject기능성 근적외선 뇌 영상장치 (fNIRS)-
dc.subject뇌 광학 이미징-
dc.subject헤모다이나믹스-
dc.subject정합필터-
dc.subject높은 신호 대 잡음비-
dc.subject변화량 대 잡음비-
dc.subject코드분할다중접속-
dc.subject출력 파워 컨트롤-
dc.subject휴대용-
dc.subject실시간-
dc.titleMonolithic expandable IC based portable fNIRS system design-
dc.title.alternative확장이 가능한 단일 집적회로 기반의 휴대용 근적외선 뇌 영상장치 개발-
dc.typeThesis(Ph.D)-
dc.identifier.CNRN325007-
dc.description.department한국과학기술원 :전기및전자공학부,-
Appears in Collection
EE-Theses_Ph.D.(박사논문)
Files in This Item
There are no files associated with this item.