Case study on human alpha 1-antitrypsin: Recombinant protein titers obtained by commercial ELISA kits are inaccurate

Cited 5 time in webofscience Cited 0 time in scopus
  • Hit : 530
  • Download : 0
DC FieldValueLanguage
dc.contributor.authorHansen, Henning Gramko
dc.contributor.authorKildegaard, Helene Faustrupko
dc.contributor.authorLee, Gyun Minko
dc.contributor.authorKol, Stefanko
dc.date.accessioned2017-01-18T02:40:44Z-
dc.date.available2017-01-18T02:40:44Z-
dc.date.created2017-01-04-
dc.date.created2017-01-04-
dc.date.created2017-01-04-
dc.date.issued2016-12-
dc.identifier.citationBIOTECHNOLOGY JOURNAL, v.11, no.12, pp.1648 - 1656-
dc.identifier.issn1860-6768-
dc.identifier.urihttp://hdl.handle.net/10203/219629-
dc.description.abstractAccurate titer determination of recombinant proteins is crucial for evaluating protein production cell lines and processes. Even though enzyme-linked immunosorbent assay (ELISA) is the most widely used assay for determining protein titer, little is known about the accuracy of commercially available ELISA kits. We observed that estimations of recombinant human alpha 1-antitrypsin (r alpha 1AT) titer by Coomassie-stained SDS-PAGE gels did not correspond to previously obtained titers obtained by a commercially available ELISA kit. This prompted us to develop two independent quantification assays based on biolayer interferometry and reversed-phase high-performance liquid chromatography. We compared the r alpha 1AT titer obtained by these assays with three different off-the-shelf ELISA kits and found that the ELISA kits led to inconsistent results. The data presented here show that recombinant protein titers determined by ELISA kits cannot be trusted per se. Consequently, any ELISA kit to be used for determining recombinant protein titer must be validated by a different, preferably orthogonal method.-
dc.languageEnglish-
dc.publisherWILEY-V C H VERLAG GMBH-
dc.subjectHAMSTER OVARY CELLS-
dc.subjectOPTICAL BIOSENSOR-
dc.subjectN-GLYCOSYLATION-
dc.subjectEXPRESSION-
dc.subjectALPHA(1)-ANTITRYPSIN-
dc.subjectTECHNOLOGIES-
dc.subjectPRODUCTS-
dc.subjectASSAY-
dc.subjectLINE-
dc.titleCase study on human alpha 1-antitrypsin: Recombinant protein titers obtained by commercial ELISA kits are inaccurate-
dc.typeArticle-
dc.identifier.wosid000390700800017-
dc.identifier.scopusid2-s2.0-85001908127-
dc.type.rimsART-
dc.citation.volume11-
dc.citation.issue12-
dc.citation.beginningpage1648-
dc.citation.endingpage1656-
dc.citation.publicationnameBIOTECHNOLOGY JOURNAL-
dc.identifier.doi10.1002/biot.201600409-
dc.contributor.localauthorLee, Gyun Min-
dc.contributor.nonIdAuthorHansen, Henning Gram-
dc.contributor.nonIdAuthorKildegaard, Helene Faustrup-
dc.contributor.nonIdAuthorKol, Stefan-
dc.description.isOpenAccessN-
dc.type.journalArticleArticle-
dc.subject.keywordAuthorAlpha-1 antitrypsin-
dc.subject.keywordAuthorBiolayer interferometry-
dc.subject.keywordAuthorChinese hamster ovary (CHO) cells-
dc.subject.keywordAuthorELISA-
dc.subject.keywordAuthorRP-HPLC-
dc.subject.keywordPlusHAMSTER OVARY CELLS-
dc.subject.keywordPlusOPTICAL BIOSENSOR-
dc.subject.keywordPlusN-GLYCOSYLATION-
dc.subject.keywordPlusEXPRESSION-
dc.subject.keywordPlusALPHA(1)-ANTITRYPSIN-
dc.subject.keywordPlusTECHNOLOGIES-
dc.subject.keywordPlusPRODUCTS-
dc.subject.keywordPlusASSAY-
dc.subject.keywordPlusLINE-
Appears in Collection
BS-Journal Papers(저널논문)
Files in This Item
There are no files associated with this item.
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 5 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0