Effects of Skin Surface Temperature on Photoplethysmograph

Cited 21 time in webofscience Cited 14 time in scopus
  • Hit : 166
  • Download : 0
DC FieldValueLanguage
dc.contributor.authorJeong, In Cheolko
dc.contributor.authorYoon, Hyungroko
dc.contributor.authorKang, Hyun Jeongko
dc.contributor.authorYeom, Hojunko
dc.date.accessioned2015-11-20T09:08:54Z-
dc.date.available2015-11-20T09:08:54Z-
dc.date.created2015-02-24-
dc.date.created2015-02-24-
dc.date.issued2014-12-
dc.identifier.citationJOURNAL OF HEALTHCARE ENGINEERING, v.5, no.4, pp.429 - 438-
dc.identifier.issn2040-2295-
dc.identifier.urihttp://hdl.handle.net/10203/201039-
dc.description.abstractPhotoplethysmograph (PPG) has been widely used to investigate various cardiovascular conditions. Previous studies demonstrated effects of temperature of the measurement environment; however, an integrated evaluation has not been established in environments with gradual air temperature variation. The purpose of this study is to investigate variations and relationships of blood pressure (BP), PPG and cardiovascular parameters such as heart rate (HR), stroke volume (SV), cardiac output (CO) and total peripheral resistance (TPR), by changing skin surface temperature (SST). Local mild cooling and heating was conducted on 16 healthy subjects. The results showed that local SST changes affected Finometer blood pressures (Finger BP), PPG components and TPR, but not the oscillometric blood pressure (Central BP), HR, SV and CO, and indicated that temperature must be maintained and monitored to reliably evaluate cardiovascular conditions in temperature-varying environments.-
dc.languageEnglish-
dc.publisherMULTI-SCIENCE PUBL CO LTD-
dc.subjectBLOOD-PRESSURE-
dc.subjectVOLUME-
dc.titleEffects of Skin Surface Temperature on Photoplethysmograph-
dc.typeArticle-
dc.identifier.wosid000347620000004-
dc.identifier.scopusid2-s2.0-84925229575-
dc.type.rimsART-
dc.citation.volume5-
dc.citation.issue4-
dc.citation.beginningpage429-
dc.citation.endingpage438-
dc.citation.publicationnameJOURNAL OF HEALTHCARE ENGINEERING-
dc.identifier.doi10.1260/2040-2295.5.4.429-
dc.contributor.nonIdAuthorJeong, In Cheol-
dc.contributor.nonIdAuthorYoon, Hyungro-
dc.contributor.nonIdAuthorYeom, Hojun-
dc.type.journalArticleArticle-
dc.subject.keywordAuthorPhotoplethysmography-
dc.subject.keywordAuthorlocal cooling and heating-
dc.subject.keywordAuthorskin surface temperature-
dc.subject.keywordAuthorcardiovascular status-
dc.subject.keywordAuthorvasoconstriction-
dc.subject.keywordAuthorvasodilation-
dc.subject.keywordPlusBLOOD-PRESSURE-
dc.subject.keywordPlusVOLUME-