Test of prototype ITER vacuum ultraviolet spectrometer and its application to impurity study in KSTAR plasmas

Cited 14 time in webofscience Cited 13 time in scopus
  • Hit : 391
  • Download : 0
DC FieldValueLanguage
dc.contributor.authorSeon, C. R.ko
dc.contributor.authorHong, J. H.ko
dc.contributor.authorJang, Juhyeokko
dc.contributor.authorLee, S. H.ko
dc.contributor.authorChoe, Won-Hoko
dc.contributor.authorLee, H. H.ko
dc.contributor.authorCheon, M. S.ko
dc.contributor.authorPak, S.ko
dc.contributor.authorLee, H. G.ko
dc.contributor.authorBiel, W.ko
dc.contributor.authorBarnsley, R.ko
dc.date.accessioned2015-04-07T02:47:18Z-
dc.date.available2015-04-07T02:47:18Z-
dc.date.created2014-12-29-
dc.date.created2014-12-29-
dc.date.created2014-12-29-
dc.date.issued2014-11-
dc.identifier.citationREVIEW OF SCIENTIFIC INSTRUMENTS, v.85, no.11-
dc.identifier.issn0034-6748-
dc.identifier.urihttp://hdl.handle.net/10203/195142-
dc.description.abstractTo optimize the design of ITER vacuum ultraviolet (VUV) spectrometer, a prototype VUV spectrometer was developed. The sensitivity calibration curve of the spectrometer was calculated from the mirror reflectivity, the grating efficiency, and the detector efficiency. The calibration curve was consistent with the calibration points derived in the experiment using the calibrated hollow cathode lamp. For the application of the prototype ITER VUV spectrometer, the prototype spectrometer was installed at KSTAR, and various impurity emission lines could be measured. By analyzing about 100 shots, strong positive correlation between the O VI and the C IV emission intensities could be found.-
dc.languageEnglish-
dc.publisherAMER INST PHYSICS-
dc.titleTest of prototype ITER vacuum ultraviolet spectrometer and its application to impurity study in KSTAR plasmas-
dc.typeArticle-
dc.identifier.wosid000345646000151-
dc.identifier.scopusid2-s2.0-84904207298-
dc.type.rimsART-
dc.citation.volume85-
dc.citation.issue11-
dc.citation.publicationnameREVIEW OF SCIENTIFIC INSTRUMENTS-
dc.identifier.doi10.1063/1.4886430-
dc.contributor.localauthorChoe, Won-Ho-
dc.contributor.nonIdAuthorSeon, C. R.-
dc.contributor.nonIdAuthorHong, J. H.-
dc.contributor.nonIdAuthorLee, S. H.-
dc.contributor.nonIdAuthorLee, H. H.-
dc.contributor.nonIdAuthorCheon, M. S.-
dc.contributor.nonIdAuthorPak, S.-
dc.contributor.nonIdAuthorLee, H. G.-
dc.contributor.nonIdAuthorBiel, W.-
dc.contributor.nonIdAuthorBarnsley, R.-
dc.description.isOpenAccessN-
dc.type.journalArticleArticle-
dc.subject.keywordPlusCURRENT HOLLOW-CATHODE-
dc.subject.keywordPlusWAVELENGTH RANGE-
dc.subject.keywordPlusDETECTOR-
dc.subject.keywordPlusSTANDARD-
dc.subject.keywordPlusREGION-
Appears in Collection
NE-Journal Papers(저널논문)
Files in This Item
There are no files associated with this item.
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 14 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0