DC Field | Value | Language |
---|---|---|
dc.contributor.author | Kim, Chang-Hoon | ko |
dc.contributor.author | Ahn, Jae-Hyuk | ko |
dc.contributor.author | Lee, Kyung-Bok | ko |
dc.contributor.author | Jung, Cheul-Hee | ko |
dc.contributor.author | Park, Hyun-Gyu | ko |
dc.contributor.author | Choi, Yang-Kyu | ko |
dc.date.accessioned | 2013-03-12T17:03:58Z | - |
dc.date.available | 2013-03-12T17:03:58Z | - |
dc.date.created | 2012-10-30 | - |
dc.date.created | 2012-10-30 | - |
dc.date.created | 2012-10-30 | - |
dc.date.issued | 2012-10 | - |
dc.identifier.citation | IEEE TRANSACTIONS ON ELECTRON DEVICES, v.59, no.10, pp.2825 - 2831 | - |
dc.identifier.issn | 0018-9383 | - |
dc.identifier.uri | http://hdl.handle.net/10203/102950 | - |
dc.description.abstract | A new sensing metric is proposed for a field-effect transistor (FET)-based biosensor. As proof of concept, a nanogap-embedded FET is studied to reduce data fluctuations that originate from process variations during FET fabrication and environmental variations stemming from bioexperiments. The new sensing metric utilizes a crucial gate voltage (V-G@I-sub,I-max), which induces the maximum substrate current. The new sensing metric shows higher immunity against variations of the nanogap length, compared with the commonly used metric that relies on threshold voltage or drain current. The proposed metric also shows smaller fluctuation, which is caused by environmental variation coming from biotreatment steps. This analysis is verified experimentally and proved by device simulations. For simple analysis, the effect of external charge of the biomolecules is eliminated by using peptide nucleic acid, which is an electrically neutral biomolecule. Thus, by using such biomolecules, the permittivity effect rising from the biomolecules within the nanogap of the gate dielectric is investigated. | - |
dc.language | English | - |
dc.publisher | IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC | - |
dc.title | A New Sensing Metric to Reduce Data Fluctuations in a Nanogap-Embedded Field-Effect Transistor Biosensor | - |
dc.type | Article | - |
dc.identifier.wosid | 000309132200037 | - |
dc.identifier.scopusid | 2-s2.0-84866741426 | - |
dc.type.rims | ART | - |
dc.citation.volume | 59 | - |
dc.citation.issue | 10 | - |
dc.citation.beginningpage | 2825 | - |
dc.citation.endingpage | 2831 | - |
dc.citation.publicationname | IEEE TRANSACTIONS ON ELECTRON DEVICES | - |
dc.identifier.doi | 10.1109/TED.2012.2209650 | - |
dc.embargo.liftdate | 9999-12-31 | - |
dc.embargo.terms | 9999-12-31 | - |
dc.contributor.localauthor | Park, Hyun-Gyu | - |
dc.contributor.localauthor | Choi, Yang-Kyu | - |
dc.contributor.nonIdAuthor | Lee, Kyung-Bok | - |
dc.description.isOpenAccess | N | - |
dc.type.journalArticle | Article | - |
dc.subject.keywordAuthor | Coefficient of variation (CV) | - |
dc.subject.keywordAuthor | drain current | - |
dc.subject.keywordAuthor | field-effect transistor (FET)-based biosensor | - |
dc.subject.keywordAuthor | impact ionization | - |
dc.subject.keywordAuthor | label free | - |
dc.subject.keywordAuthor | nanogap | - |
dc.subject.keywordAuthor | nanogap-embedded FET | - |
dc.subject.keywordAuthor | peptide nucleic acid (PNA) | - |
dc.subject.keywordAuthor | reliability | - |
dc.subject.keywordAuthor | substrate current | - |
dc.subject.keywordAuthor | threshold voltage | - |
dc.subject.keywordPlus | SILICON NANOWIRE BIOSENSORS | - |
dc.subject.keywordPlus | ELECTRICAL DETECTION | - |
dc.subject.keywordPlus | HYBRIDIZATION | - |
dc.subject.keywordPlus | SENSORS | - |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.