Use of a flat-panel detector for microtomography: A feasibility study for small-animal imaging

Cited 29 time in webofscience Cited 30 time in scopus
  • Hit : 2044
  • Download : 1528
We have applied a flat-panel detector to an X-ray cone-beam micro computed tomography (micro-CT) for small-animal imaging. The flat-panel detector consists of an active matrix of transistors and photodiodes with a pixel pitch of 50 mu m and a thallium-doped cesium iodide (CsI:Tl) scintillator as an X-ray-to-light conversion layer. The detector was fabricated with a complementary metal-oxide-semiconductor (CMOS) technology capable of a submicrometer design line width, hence, it has a pixel fill-factor as high as similar to 80%. In addition, the detector has a very fast response characteristic with an image lag less than 0.3% in the frame integration time of 5 s. The CMOS flat-panel detector has been tested in terms of modulation transfer function, noise power spectrum, and detective quantum efficiency. Tomographic imaging performances of the micro-CT system, such as voxel noise, contrast-to-noise ratio, and spatial resolution, have also been evaluated by using various quantitative phantoms. Experimental results of euthanized laboratory rat imaging suggest that the micro-CT system employing a CMOS flat-panel detector can be greatly used in small-animal imaging.
Publisher
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
Issue Date
2005-02
Language
English
Article Type
Article
Keywords

BEAM COMPUTED-TOMOGRAPHY; IMAGER; PERFORMANCE

Citation

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, v.52, no.1, pp.193 - 198

ISSN
0018-9499
DOI
10.1109/TNS.2004.843107
URI
http://hdl.handle.net/10203/10284
Appears in Collection
NE-Journal Papers(저널논문)
Files in This Item
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 29 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0