Deconvolution of gamma-ray spectra obtained with NAI(Tl) detector in a water tank

Cited 7 time in webofscience Cited 0 time in scopus
  • Hit : 680
  • Download : 0
Maximum-likelihood fitting by the expectation maximisation deconvolution method is presented to analyse gamma-ray spectra recorded using an NaI(Tl) detector for a water monitoring system. The applicability of the method was tested by deconvolving measured spectra taken using an industry standard 3 ' x 3 ' cylindrical NaI(Tl) detector in a model water tank with several calibration sources. The results show significant removal of the Compton continuum counts and efficient transfer of the counts into the corresponding photo-peaks. The peak-to-total count ratio and the number of counts in the photo-peaks in the deconvolved spectra increased approximately 4.67 and 5.29 times, respectively, compared with those of measured spectra taken using an NaI(Tl) scintillation detector in the case of (137)Cs.
Publisher
OXFORD UNIV PRESS
Issue Date
2009-07
Language
English
Article Type
Article
Keywords

SCINTILLATION DETECTORS; MAXIMUM-LIKELIHOOD; RESPONSE MATRIX; RADIOACTIVITY; STATISTICS; SIMULATION

Citation

RADIATION PROTECTION DOSIMETRY, v.135, no.3, pp.203 - 210

ISSN
0144-8420
DOI
10.1093/rpd/ncp102
URI
http://hdl.handle.net/10203/101139
Appears in Collection
NE-Journal Papers(저널논문)