레이저 간섭계의 진직도 측정오차 보상Compensation of the Straightness Measurement Error in the Laser Interferometer

Cited 0 time in webofscience Cited 0 time in scopus
  • Hit : 399
  • Download : 0
DC FieldValueLanguage
dc.contributor.author김경호ko
dc.contributor.author김태호ko
dc.contributor.author이후상ko
dc.contributor.author김승우ko
dc.date.accessioned2013-03-06T12:41:44Z-
dc.date.available2013-03-06T12:41:44Z-
dc.date.created2012-02-06-
dc.date.created2012-02-06-
dc.date.issued2005-09-
dc.identifier.citation한국정밀공학회지, v.22, no.9, pp.69 - 76-
dc.identifier.issn1225-9071-
dc.identifier.urihttp://hdl.handle.net/10203/87008-
dc.description.abstractThe laser interferometer system such as HP5529A is one of the most powerful equipment fur measurement of the straightness error in precision stages. The straightness measurement system, HP5529A is composed of a Wollaston prism and a reflector. In this system, the straightness error is defined as relative lateral motion change between the prism and the reflector and computed from optical path difference of two polarized laser beams between these optics. However, rotating motion of the prism or the reflector used as a moving optic causes unwanted straightness error. In this paper, a compensation method is proposed for removing the unwanted straightness error generated by rotating the moving optic and an experiment is carried out for theoretical verification. The result shows that the unwanted straightness error becomes very large when the reflector is used as the moving optic and the distance between the reflector and the prism is far. Therefore, the prism must be generally used as the moving optic instead of the reflector so as to reduce the measurement error. Nevertheless, the measurement error must be compensated because its not a negligible error if a rotating angle of the prism is large. In case the reflector must be used as the moving optic, which is unavoidable when the squareness error is measured between two axes, this compensation method can be applied and produces a better result.-
dc.languageKorean-
dc.publisher한국정밀공학회-
dc.title레이저 간섭계의 진직도 측정오차 보상-
dc.title.alternativeCompensation of the Straightness Measurement Error in the Laser Interferometer-
dc.typeArticle-
dc.type.rimsART-
dc.citation.volume22-
dc.citation.issue9-
dc.citation.beginningpage69-
dc.citation.endingpage76-
dc.citation.publicationname한국정밀공학회지-
dc.identifier.kciidART001190342-
dc.contributor.localauthor김승우-
dc.contributor.nonIdAuthor김경호-
dc.contributor.nonIdAuthor김태호-
dc.contributor.nonIdAuthor이후상-
dc.subject.keywordAuthorLaser interferometer-
dc.subject.keywordAuthorStraightness measurement-
dc.subject.keywordAuthorAngular error-
dc.subject.keywordAuthorError compensation-
dc.subject.keywordAuthorPrecision stage-
dc.subject.keywordAuthor레이저 간섭계-
dc.subject.keywordAuthor진직도 측정-
dc.subject.keywordAuthor각도 오차-
dc.subject.keywordAuthor오차 보상-
dc.subject.keywordAuthor정밀 스테이지-
dc.subject.keywordAuthorLaser interferometer-
dc.subject.keywordAuthorStraightness measurement-
dc.subject.keywordAuthorAngular error-
dc.subject.keywordAuthorError compensation-
dc.subject.keywordAuthorPrecision stage-