Forward projection matrix derivation through Monte-Carlo ray-tracing of KSTAR infra-red imaging video bolometer (IRVB)

Cited 0 time in webofscience Cited 0 time in scopus
  • Hit : 425
  • Download : 0
The infrared imaging video bolometer (IRVB) as a foil bolometry technique can be an alternative solution to the conventional resistive bolometer due to its electro-magnetic immunity and 2D plasma profiles. The plasma profile of the IRVB cannot be directly derived from the foil images due to the difference between the foil pixel number and the plasma pixel number and the line integrated nature of the incident rays on the foil. So, it needs tomography such as the Phillips-Thikhonov algorithm. The projection matrix constructing the foil image from the plasma very directly influences the tomography. So, the projection matrix needs to be constructed precisely. For the technique calculating the precise projection matrix, a forward Monte-Carlo ray-tracing method is proposed here, and this technique can provide the detailed descriptions of the foil image. And it can give enhanced performance in the reconstructions of the plasmas with spatially localized power. Published by AIP Publishing.
Publisher
AMER INST PHYSICS
Issue Date
2018-10
Language
English
Article Type
Article; Proceedings Paper
Citation

REVIEW OF SCIENTIFIC INSTRUMENTS, v.89, no.10, pp.10E118

ISSN
0034-6748
DOI
10.1063/1.5036929
URI
http://hdl.handle.net/10203/247136
Appears in Collection
NE-Journal Papers(저널논문)
Files in This Item
There are no files associated with this item.

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0