A Feasibility Study on the Adoption of Human Body Communication for Medical Service

Cited 17 time in webofscience Cited 17 time in scopus
  • Hit : 296
  • Download : 0
A feasibility study on the adoption of human body communication (HBC) for medical service is presented. The electrical specifications of HBC for medical service were reviewed. The ratio of the leakage current into the heart to the output current of a capsule-type transmitter and the propagation loss in the channel were simulated using an anatomical body model. The in-body to on-body channel loss for HBC was provided by electromagnetic simulation from the capsule-type transmitter in the stomach to receiving electrodes on the abdominal region. An analog front end with a sensitivity of 37 mu Vpp was developed, which showed an improvement of 21 dB against the best report. The 14 mm x 26 mm capsule-type transmitter was implemented, and preclinical experiment was conducted using a live pig. The proposed system successfully transferred data from the capsule-type transmitter in a stomach to the electrodes on the butt and the low jaw of the pig at a high rate of 2 Mb/s at a distance of 90 cmwithout packet error.
Publisher
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
Issue Date
2015-02
Language
English
Article Type
Article
Keywords

CHANNEL COMMUNICATION; TRANSCEIVER

Citation

IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS II-EXPRESS BRIEFS, v.62, pp.169 - 173

ISSN
1549-7747
DOI
10.1109/TCSII.2014.2387631
URI
http://hdl.handle.net/10203/195786
Appears in Collection
EE-Journal Papers(저널논문)
Files in This Item
There are no files associated with this item.
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 17 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0