Development of a closed-loop J-T cryoablation device with a long cooling area and multiple expansion parts

Cited 8 time in webofscience Cited 9 time in scopus
  • Hit : 346
  • Download : 0
DC FieldValueLanguage
dc.contributor.authorLee, Cheonkyuko
dc.contributor.authorBaek, Seung-Whanko
dc.contributor.authorLee, Jisungko
dc.contributor.authorHwang, Gyuwanko
dc.contributor.authorJeong, Sang-Kwonko
dc.contributor.authorPark, Sang Wooko
dc.date.accessioned2015-01-29T07:29:17Z-
dc.date.available2015-01-29T07:29:17Z-
dc.date.created2014-09-05-
dc.date.created2014-09-05-
dc.date.created2014-09-05-
dc.date.created2014-09-05-
dc.date.issued2014-11-
dc.identifier.citationMEDICAL ENGINEERING PHYSICS, v.36, no.11, pp.1464 - 1472-
dc.identifier.issn1350-4533-
dc.identifier.urihttp://hdl.handle.net/10203/193913-
dc.description.abstractCryoablation is a surgical procedure used to freeze defective cells by inserting a low temperature probe into a human body to destroy malignant tissues. Miniaturized Joule-Thomson (J-T) refrigerators are often used to minimize the volume of the cooling device and reduce the destruction zone of normal tissue. The cooling effects of the existing probes are not uniformly generated along the longitudinal direction of the probe, which makes their applications less effective in surgeries of incompetent great saphenous veins (GSVs), where the target cells are distributed over a broad range. Long uniform refrigeration is required across the entire area of the probe to apply the same cooling effects. In this paper, a closed-loop J-T cryoablation probe was designed and fabricated to provide uniform refrigeration over a large area, with multiple expansion parts. Using flow boiling heat transfer, uniform cooling of a 200 mm-long and 0.3 mm thickness piece of target tissue was possible and simulated in a gelatin solution. The developed probe produced a greater than 53 K min(-1) cooling rate and the cooling temperature was below 253 K to satisfy the required cell death conditions.-
dc.languageEnglish-
dc.publisherELSEVIER SCI LTD-
dc.subjectCRYOSURGERY-
dc.titleDevelopment of a closed-loop J-T cryoablation device with a long cooling area and multiple expansion parts-
dc.typeArticle-
dc.identifier.wosid000345108100014-
dc.identifier.scopusid2-s2.0-84922641355-
dc.type.rimsART-
dc.citation.volume36-
dc.citation.issue11-
dc.citation.beginningpage1464-
dc.citation.endingpage1472-
dc.citation.publicationnameMEDICAL ENGINEERING PHYSICS-
dc.identifier.doi10.1016/j.medengphy.2014.07.017-
dc.contributor.localauthorJeong, Sang-Kwon-
dc.contributor.nonIdAuthorLee, Jisung-
dc.contributor.nonIdAuthorHwang, Gyuwan-
dc.contributor.nonIdAuthorPark, Sang Woo-
dc.type.journalArticleArticle-
dc.subject.keywordAuthorCryoablation probe-
dc.subject.keywordAuthorJ-T refrigerator-
dc.subject.keywordAuthorMultiple-expansion-
dc.subject.keywordAuthorUniform cooling-
dc.subject.keywordAuthorClosed system-
dc.subject.keywordAuthorCryoablation probe-
dc.subject.keywordAuthorJ-T refrigerator-
dc.subject.keywordAuthorMultiple-expansion-
dc.subject.keywordAuthorUniform cooling-
dc.subject.keywordAuthorClosed system-
dc.subject.keywordPlusCRYOSURGERY-
dc.subject.keywordPlusCRYOSURGERY-
Appears in Collection
ME-Journal Papers(저널논문)
Files in This Item
There are no files associated with this item.
This item is cited by other documents in WoS
⊙ Detail Information in WoSⓡ Click to see webofscience_button
⊙ Cited 8 items in WoS Click to see citing articles in records_button

qr_code

  • mendeley

    citeulike


rss_1.0 rss_2.0 atom_1.0