교차 전기영동법을 이용한 극소형 DNA 추출기A DNA Microextractor Using Crossed Field Electrophoresis

Cited 0 time in webofscience Cited 0 time in scopus
  • Hit : 483
  • Download : 0
This paper presents a microextractor for the separation of DNA molecules by their sizes. The DNA microextractor immobilizes the DNA molecules of specific size in the micropillar array by adjusting the period of the crossed electric field, thus providing a starting-point independent target DNA extraction method without separation process monitoring. The DNA microextractor has been fabricated by a three-mask micromachining process. The velocity of three different DNA molecules has been measured at the electric field of E=5V/0.8cm in the fabricated DNA microextractor, resulting in the reorientation times of , and for DNA, micrococcus DNA, and T4 DNA, respectively. T4 DNA is trapped in the micropillar array when the crossed electric field of 5V/0.8cm is applied alternately at a 10 second time interval. The present DNA microextractor filters the DNA in a specific size range by adjusting the magnitude and/or the period of the crossed electric field applied in the micropillar array.
Publisher
대한기계학회
Issue Date
2004-08
Language
Korean
Citation

대한기계학회논문집 A, v.28, no.8, pp.1135 - 1139

ISSN
1226-4873
URI
http://hdl.handle.net/10203/84344
Appears in Collection
BiS-Journal Papers(저널논문)