개량된 다채널 임피던스형 측정기에 의한 기포율의 측정Void Fraction Measurement by the Improved Multi-Channel Impedance Void Meter

Cited 0 time in webofscience Cited 0 time in scopus
  • Hit : 494
  • Download : 0
An improved multi-channel Impedance Void Meter (IVM) is developed to measure an area-averaged void fraction. It consists of a main sensor, a reference sensor and a signal processor. The sensor was designed to be flush-mounted to the inner wall of the test section to avoid the flow disturbances. Guard electrodes are used to obtain evenly distributed electrical field in a measuring volume. A reference sensor is also installed to eliminate the drift in void signal caused by the changes in electrical properties of working fluid. The signal processor with three channels is specially designed so as to minimize the inherent error due to the phase difference between channels. As an example of applications, the mean and fluctuating components of void fraction are measured for bubbly and slug flow regime, and it is shown that IVM has good dynamic resolution which is required to investigate the structural developments of bubbly flow and the propagation of void waves in a flow channel.
Publisher
대한기계학회
Issue Date
1996-01
Language
Korean
Citation

대한기계학회논문집 B, v.20, no.1, pp.384 - 398