Measurement of the Catheter Motion for Catheter Simulation

Cited 0 time in webofscience Cited 0 time in scopus
  • Hit : 125
  • Download : 0
This paper presents a method to measure the catheter motion for a simulation of medical catheterization procedures. An optical navigation sensor used in a common computer mouse is utilized to measure the translational and rotational velocity of the clinical catheter. Catheter passage is designed to keep the catheter inside the effective sensing area of the optical sensor. Experiment results show that the developed device can measure the velocity up to 150 mm/s in translation and 6 rad/s in rotation. Errors are 6.78 % and 19.5 % in the axial and radial measurements, respectively. The RMS errors are 3.69 mm/s and 0.6 rad/s for the respective directions.
Publisher
ICCAS
Issue Date
2021-10-14
Language
English
Citation

2021 The 21st International Conference on Control, Automation and Systems (ICCAS 2021)

URI
http://hdl.handle.net/10203/288548
Appears in Collection
ME-Conference Papers(학술회의논문)
Files in This Item
There are no files associated with this item.

qr_code

  • mendeley

    citeulike