COMPACT OCT ENDOMICROSCOPIC CATHETER USING FLIP-CHIP BONDED LISSAJOUS SCANNED ELECTROTHERMAL MEMS FIBER SCANNER

Cited 6 time in webofscience Cited 0 time in scopus
  • Hit : 201
  • Download : 0
We report a compact optical coherence tomography (OCT) endomicroscope with a Lissajous scanned electrothermal MEMS fiber scanner. Two-dimensional Lissajous scanning was realized by Joule heating of the electrothermal MEMS fiber scanner. The compact packaging of MEMS scanner has a primary challenge, particularly in inserting it into an endomicroscopic catheter. The electrothermal MEMS fiber scanner was precisely designed and flip-chip bonded with a thin printed circuit board and completely packaged with 1.65 mm diameter housing tube, 1 mm diameter GRIN lens, and a single mode optical fiber. A 1.65 mm diameter OCT endomicroscopic catheter was successfully combined with a spectral-domain OCT system. Two-dimensional OCT image of a finger nail was successfully obtained with the endomicroscopic catheter.
Publisher
MEMS association
Issue Date
2017-01-24
Language
English
Citation

30th IEEE International Conference on Micro Electro Mechanical Systems (MEMS), pp.518 - 521

ISSN
1084-6999
DOI
10.1109/MEMSYS.2017.7863457
URI
http://hdl.handle.net/10203/237639