A nonresonant, fiber-optic raster scanning endomicroscope was developed using a quarter-tubular piezoelectric (PZT) actuator. A fiber lever mechanism was utilized to enhance the small actuation range of the tubular PZT actuator and to increase its field-of-view. Finite element method simulation of the endoscopic probe was conducted for various conditions to maximize its scanning range. After fabricating the probe using a double clad fiber, we obtained two-photon fluorescence images using raster beam scanning of the fiber. The outer diameter of the probe was 3.5 mm and its rigid distal length was 30 mm including a high numerical aperture gradient index lens. These features are sufficient for input into the instrumental channel of a commercial colonoscope or gastroscope to obtain high resolution images in vivo. (C) The Authors. Published by SPIE under a Creative Commons Attribution 3.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.